May 17, 2012  |  Home  |  Vasomedical Store  |  VasoHealthcare  |  EECP Forum  |  Contact
tvl
What's New
Vasomedical-BIOX™ 1303 Ultra Compact 3-Channel ECG Holter Recorder

FDA CLEARED/CE MARKED
ECG Holter Recorder

Vasomedical-BIOX™ 1303
Ultra Compact 3-Channel ECG Holter Recorder

Vasomedical Biox 2301

Vasomedical-BIOX™ Model 1303

The Vasomedical-BIOX™ 1303 Ultra Compact 3-Channel ECG Holter recorder with LCD screen captures all the important data you require,
while allowing your patient the ease and comfort of one of the smallest ECG Holter recorders on the market today.

Click Here To Buy Model 1303 Online! >

Contact Us About This Product >


Features
Advanced Technology:
  • Advanced miniaturization provides an ultra compact, ultra lightweight ECG Holter design (2.6 x 2.1 x 0.7 inches3 / 65 x 52 x 18 mm3 and 1.76 oz / 50 g excluding AAA battery)
  • High resolution and sample rate enable high quality ECG signal capture with less artifact
  • Advanced amplifier and A/D converter design allows for exceptional signal fidelity
  • Independent pacemaker and VLP detection
  • Low power consumption - One AAA battery allows recording for 72 hours

Convenience and ease of use:

  • Easy-to-read LCD display enables real time ECG viewing hook-up, eliminating inefficient and
    costly re-testing
  • Easy push-button setup and fast download/analysis reduce technician time
  • Removable SD memory card enables quick data transfer and turnaround of the recorder
  • Ultra compact and lightweight for patient comfort
Specifications
GENERAL
Accessories: • 2 GB SD memory card
• USB SD memory card reader
• 5 or 7 lead ECG patient cable
• Carrying case & strap
• Recorder user manual
• User Software CD

Recording time:

24, 48 or 72 hours

Storage medium:

SD card

Memory size:

1 GB or more

Data transfer:

Full disclosure transfer via USB card reader

Batteries:

1 “AAA” size alkaline batteries

Dimensions:

2.6 x 2.1 x 0.7 (inches3) / 65 x 52 x 18 (mm3)

Weight:

1.76 oz / 50 g (without battery)

Operation Environment: Temperature: 50 - 104ºF / 10 - 45º C
Relative humidity: 10 - 95%
Atmospheric pressure: 646.6 - 797mmHg / 86 - 106 kPa

Regulatory compliance:

FDA cleared / CE Marked / Health Canada Listed
ISO 13485 Certified

ECG

No. of ECG Channels:

3

No. of ECG electrodes:

5 or 7

Pacemaker detection:

Specified during setup (using software)

Input dynamic range: +5 mV ECG, +300 mV offset voltage
Input impedance: >10 MΩ
CMMR: >80 dB

Gain accuracy:

Maximum amplitude error <10%

Gain stability: Change <3% over a 24-hour period
System noise: <50 μVp-v
Multi-channel crosstalk: <0.2 mVp-v
Bandwidth: 0.05 Hz - 40 Hz (±3 dB)
Minimum feature size: 50 μVp-v
Timing accuracy: Overall error during 24-hour period <30 s

Sampling rate:

256 Hz/channel
10,000 Hz/channel for pacemaker detection
1,000 Hz/channel with SAECG option

Resolution:

10 bit
12 bit with SAECG option