February 4, 2012  |  Home  |  Vasomedical Store  |  VasoHealthcare  |  EECP Forum  |  Contact
tvl
What's New
Vasomedical-BIOX™ 2301, Combined ECG Holter and ABP Monitor

FDA CLEARED/CE MARKED
Synchronous ECG & ABPM Monitoring

  • 3 Modes Of Operation: ECG, ABPM
    & Combined ECG & ABPM
  • Interactive 3-Channel ECG &
    ABPM recording & storage for 24 Hours
  • Wireless infrared port for real-time ECG & ABPM verification
  • Low power consumption
  • High sampling rate and high resolution ECG recording
  • Pacemaker detection
Click Here To Buy Online! >
Contact Us About This Product >

Specifications
GENERAL
Accessories: − One 2GB SD card
− Lead wires
− Blood pressure cuff (standard adult size)
− Carrying case with strap
− Self-adhesive disposable electrodes
− User manual and software CD

Dimensions:

124 x 68 x 31 (mm); 4.88 x 2.68 x 1.22 (inches)

Weight:

180g / 6.34 oz (without batteries)

Battery:

4 “AA” size alkaline batteries

Recording time:

24 hours

Data transfer:

Full disclosure transfer via USB card reader; Setting and real-time data transfer via infrared port

Memory size:

1 GB or more

Storage medium:

SD card

Operating environment:

Temperature 10 - 45°C / 50 - 104°F; relative humidity 10-95%; atmospheric pressure 86-106 kPa

Regulatory compliance:

FDA 510k clearance, CE marked

ECG

No. of ECG Channels:

3

No. of ECG electrodes:

5 or 7

Pacemaker detection:

Independent channel

Input dynamic range: +5 mV ECG, +300 mV offset voltage
Input impedance: >10 MΩ
CMMR: >80 dB

Gain accuracy:

Maximum amplitude error <10%

Gain stability: Change <3% over a 24-hour period
System noise: <50 μVp-v
Multi-channel crosstalk: <0.2 mVp-v
Bandwidth: 0.05 Hz - 40 Hz (+ 3 dB)
Minimum feature size: 50 μVp-v
Timing accuracy: Overall error during 24-hour period < 30 s

Sample rate:

Maximum 10000 Hz

Preserve rate:

256 Hz

A/D Resolution:

12 bit

AMBULATORY BLOOD PRESSURE
Method of measurement: Oscillometric
Indicating range: 0 - 300 mmHg
Recording range: Systolic: 50 - 260 mmHg
Diastolic: 30 - 180 mmHg
Pressure precision: < 3 mmHg