May 17, 2012  |  Home  |  Vasomedical Store  |  VasoHealthcare  |  EECP Forum  |  Contact
tvl
What's New
Vasomedical-BIOX™ 2301/2301 Combination ECG Holter & ABPM

FDA CLEARED/CE MARKED
Combined ECG Holter & Ambulatory Blood Pressure Monitors

Vasomedical-BIOX™ 2301 Combined 3-Channel ECG Holter & Ambulatory Blood Pressure Monitor

Vasomedical-BIOX™ 2302 Combined 12-Channel ECG Holter & Ambulatory Blood Pressure Monitor

Vasomedical Biox 2301

Vasomedical-BIOX™ Model 2301

Click here for analysis software preview

Vasomedical Biox 2301

Vasomedical-BIOX™ Model 2302

Vasomedical is pleased to announce its new and unique synchronized ECG Holter and Ambulatory Blood Pressure Monitors in a single device! The Model 2301 and Model 2302 Combined Monitors provide advanced high resolution ECG Holter technology accompanied by a simultaneous presentation and analysis of blood pressure waveforms and digital results. Everything you need to obtain and verify the clinical data is now available in one lightweight, easy to use, multiparameter system.

Click Here To Buy Model 2301 Online! >
Click Here To Buy Model 2302 Online! >

Contact Us About This Product >


Features
Advanced Technology:
  • Three programmable operation modes:
    • Synchronized ECG Holter and ABPM,
    • 3 or 12-channel ECG Holter, or
    • Ambulatory blood pressure monitoring (ABPM) only
  • High resolution and sample rate enables high quality ECG signal capture with less artifact
  • Interactive recording for additional blood pressure measurements in the event of ECG abnormalities
  • Arterial pulse waveform and tabulated numerical data can be reviewed in an easy, downloadable
    format
  • Advanced amplifier and A/D converter design allows for exceptional signal fidelity
  • Dual sensor, dual MCU and dual valve technology ensure patient safety including intelligent
    overpressure protection and failsafe monitoring
  • Digital dual valve provides smooth deflation and the accuracy expected from the ausculatory method

Convenience and ease of use:

  • LCD displays measurement results as well as user-friendly icons to simplify task performance
    and data retrieval
  • Easy push-button setup and fast download/analysis reduce technician time
  • Removable SD memory card enables quick data transfer and turnaround of the device
  • Both models are small and lightweight, resulting in minimum interference with patient activity
Specifications
GENERAL
Accessories: • 2 GB SD memory card
• USB SD memory card reader
• 5 or 7 lead ECG patient cable (Model 2301)
• 10 lead ECG patient cable (Model 2302)
• Carrying case & strap
• Blood pressure cuff (Standard adult)
• Infrared communication adapter
• Recorder user manual
• User Software CD

Recording time:

24 hours

Storage medium:

SD card

Memory size:

1 GB or more

Data transfer:

Full disclosure transfer via USB card reader; Setting and real-time data transfer via infrared port

Batteries:

4 “AA” size alkaline batteries

Dimensions:

4.88 x 2.68 x 1.22 (inches3) / 124 x 68 x 31 (mm3)

Weight:

6.34 oz / 180 g (without batteries)

Regulatory compliance:

FDA cleared / CE Marked / Health Canada Listed
ISO 13485 Certified

ECG
Model 2301 Model 2302

No. of ECG Channels:

3

12

No. of ECG electrodes:

5 or 7

10

Pacemaker detection:

Independent channel

Input dynamic range: +5 mV ECG, +300 mV offset voltage
Input impedance: >10 MΩ
CMMR: >80 dB

Gain accuracy:

Maximum amplitude error <10%

Gain stability: Change <3% over a 24-hour period
System noise: <50 μVp-v
Multi-channel crosstalk: <0.2 mVp-v
Bandwidth: 0.05 Hz - 40 Hz (±3 dB)
Minimum feature size: 50 μVp-v
Timing accuracy: Overall error during 24-hour period <30 s

Sampling rate:

256 Hz/channel
2048 Hz/channel (pacemaker detection)

Resolution:

12 bit

AMBULATORY BLOOD PRESSURE
Method of measurement: Oscillometric
Indicating range: 0 - 300 mmHg
Recording range: Systolic: 50 - 260 mmHg
Diastolic: 30 - 180 mmHg
Pressure precision: <3 mmHg